1037667

Avotec, Inc. — FEI 1000113348

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
January 12, 2018
Fiscal Year
2018
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Stuart, FL (United States)

Citations

IDCFRDescription
1472221 CFR 820.40Procedures not adequately established or maintained
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
313221 CFR 820.120Lack of or inadequate procedures for labeling
323321 CFR 820.72(a)Calibration, Inspection, etc. Procedures Lack of or Inadequ
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
340921 CFR 820.200(d)Service reports
48621 CFR 820.50(a)Evaluation of suppliers, contractors, etc., requirements
63121 CFR 803.17(a)(1)Lack of System for Event Evaluations