1198080
AADCO Medical Inc. — FEI 3001237319
Inspection Details
- Classification
- Voluntary Action Indicated (VAI) (VAI)
- End Date
- February 22, 2023
- Fiscal Year
- 2023
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Randolph, VT (United States)
Citations
| ID | CFR | Description |
| 2604 | 21 CFR 820.30(e) | Design review - Lack of or inadequate procedures |
| 3104 | 21 CFR 820.30(j) | Design history file |
| 3160 | 21 CFR 820.184 | Lack of or inadequate DHR procedures |
| 3266 | 21 CFR 820.86 | Acceptance status |
| 3432 | 21 CFR 820.75(b)(2) | Documentation of validated process performance |