1198485

New Wave Endo-Surgical, Corp. — FEI 3015209038

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
February 17, 2023
Fiscal Year
2023
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Coconut Creek, FL (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
255721 CFR 820.30(c)Design input - documentation
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
421221 CFR 806.20(b)(4)Justification for not reporting
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures