1292268

3D Diagnostix Inc. — FEI 3011429437

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
November 21, 2025
Fiscal Year
2026
Product Type
Devices
Project Area
Compliance: Devices
Location
Allston, MA (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
310421 CFR 820.30(j)Design history file
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
316021 CFR 820.184Lack of or inadequate DHR procedures
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
54121 CFR 820.70(c)Environmental control Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation