705822

AGFA Healthcare Corporation — FEI 3000204644

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
January 6, 2011
Fiscal Year
2011
Product Type
Devices
Project Area
Compliance: Devices
Location
Canton, MA (United States)

Citations

IDCFRDescription
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
317321 CFR 820.198(d)Evaluation, timeliness, identification
334621 CFR 820.200(b)Analyzing service report
367721 CFR 820.30(g)Design validation - software validation not performed
419321 CFR 806.10(b)Time to report - 10 days
81221 CFR 803.56Submission of Supplement to Individual Report Within One Month