734128

Access Bio, Inc. — FEI 3008474767

Inspection Details

Classification
No Action Indicated (NAI) (NAI)
End Date
August 29, 2011
Fiscal Year
2011
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Somerset, NJ (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
232821 CFR 820.22Quality audits - auditor independence
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
237121 CFR 820.30(a)Design control - no procedures
310821 CFR 820.70(e)Contamination control, Lack of or inadequate procedures
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
315921 CFR 820.184DHR content
319221 CFR 820.30(g)Design validation - user needs and intended uses
323321 CFR 820.72(a)Calibration, Inspection, etc. Procedures Lack of or Inadequ
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
333121 CFR 820.181DMR - not or inadequately maintained
368321 CFR 820.70(g)Equipment Installation, Placement, Specified Requirements
369921 CFR 820.160(b)Distribution records
44721 CFR 820.40Lack of procedures, or not maintained
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
54121 CFR 820.70(c)Environmental control Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation