Z-0006-2023 Class II Ongoing

Recalled by Wright Medical Technology, Inc. — Memphis, TN

FDA device recall Z-0006-2023 was initiated by Wright Medical Technology, Inc. on Aug 3, 2022 and is designated Class II. Reason for recall: The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4… The recall status is ongoing. Affected quantity: 32 units.

Recall Details

Product Type
Devices
Report Date
Oct 12, 2022
Initiation Date
Aug 3, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
32 units

Product Description

stryker EVOLVE, Proline Stem, CoCr, Size: +4, OD: 9.5mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repair System is a set configuration designed to address radial head fractures. It combines parts of two systems: EVOLVE Proline and EVOLVE TRIAD.

Reason for Recall

The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm +2 stem.

Distribution Pattern

US Nationwide distribution.

Code Information

UDI/DI: (01)008404201856775(11)210831(17)290829 Lot Number 1704255

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0006-2023
openFDA file date
Nov 3, 2026
Loaded by RegDataLab
Nov 4, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0006-2023. Report a data issue: hello@regdatalab.com.