Z-0013-2020 Class II Terminated

Recalled by Inter-Med Llc — Racine, WI

FDA device recall Z-0013-2020 was initiated by Inter-Med Llc on Jun 27, 2019 and is designated Class II. Reason for recall: The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained… The recall status is terminated (terminated Aug 10, 2021). Affected quantity: 582 units.

Recall details

Product Type
Devices
Report Date
Oct 9, 2019
Initiation Date
Jun 27, 2019
Termination Date
Aug 10, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
582 units

Product description

Vista Dental Products Sodium Hypochlorite - 3%, 10 - 12cc Pre-filled Syringes, REF 502315

Reason for recall

The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

Distribution pattern

US, South Africa, Switzerland, Poland, Italy, Singapore, Australia

Code information

Lot/Work Order Numbers: 2018-1349 2018-1587 2018-1853 20182107, UDI (01)10818207020427 20182303, UDI (01)10818207020427 20182584, UDI (01)10818207020427 20182666, UDI (01)10818207020427 20190199, UDI (01)10818207020427 20190433, UDI (01)10818207020427

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0013-2020
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Inter-Med Llc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0013-2020. Report a data issue: hello@regdatalab.com.