Z-0013-2020 Class II Terminated

Recalled by Inter-Med Llc — Racine, WI

Recall Details

Product Type
Devices
Report Date
October 9, 2019
Initiation Date
June 27, 2019
Termination Date
August 10, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
582 units

Product Description

Vista Dental Products Sodium Hypochlorite - 3%, 10 - 12cc Pre-filled Syringes, REF 502315

Reason for Recall

The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

Distribution Pattern

US, South Africa, Switzerland, Poland, Italy, Singapore, Australia

Code Information

Lot/Work Order Numbers: 2018-1349 2018-1587 2018-1853 20182107, UDI (01)10818207020427 20182303, UDI (01)10818207020427 20182584, UDI (01)10818207020427 20182666, UDI (01)10818207020427 20190199, UDI (01)10818207020427 20190433, UDI (01)10818207020427