Z-0019-2018 Class II Terminated
FDA device recall Z-0019-2018 was initiated by Philips Electronics North America Corporation on Jan 18, 2017 and is designated Class II. Reason for recall: During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the… The recall status is terminated (terminated Feb 26, 2020). Affected quantity: US- 606.
Recall details
- Product Type
- Devices
- Report Date
- Oct 25, 2017
- Initiation Date
- Jan 18, 2017
- Termination Date
- Feb 26, 2020
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- US- 606
Product description
Philips Healthcare Brilliance iCT X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason for recall
During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
Distribution pattern
USA (nationwide) Distribution
Code information
Software version 4.16
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0019-2018
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab