Z-0023-2015 Class II Terminated

Recalled by Biomedix, Inc. — Bloomington, IN

Recall Details

Product Type
Devices
Report Date
October 15, 2014
Initiation Date
August 14, 2014
Termination Date
November 13, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,928

Product Description

Biomedix SELEC-3 I.V. Administration Set. Individually sealed into a Tyvek¿ and LDPE/PET pouch. Secondary packaging includes 48 pouched units per box/case. STERILE

Reason for Recall

Customer contacted Biomedix of complaint and returned the failed set along with 86 units of unopened stock (96 units originally shipped). Biomedix performed extensive investigation and found insufficient application of glue. Recall was initiated due to the rate of the failure.

Distribution Pattern

US Distribution including the states of OH, NC, and TX

Code Information

Model/Product Number: B30-102 Lot #416130 Exp. 12/2016