FDA device recall Z-0023-2015 was initiated by Biomedix, Inc. on Aug 14, 2014 and is designated Class II. Reason for recall: Customer contacted Biomedix of complaint and returned the failed set along with 86 units of unopened stock (96 units originally shipped). Biomedix performed extensive investigation and found… The recall status is terminated (terminated Nov 13, 2014). Affected quantity: 2,928.

Recall details

Product Type
Devices
Report Date
Oct 15, 2014
Initiation Date
Aug 14, 2014
Termination Date
Nov 13, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,928

Product description

Biomedix SELEC-3 I.V. Administration Set. Individually sealed into a Tyvek¿ and LDPE/PET pouch. Secondary packaging includes 48 pouched units per box/case. STERILE

Reason for recall

Customer contacted Biomedix of complaint and returned the failed set along with 86 units of unopened stock (96 units originally shipped). Biomedix performed extensive investigation and found insufficient application of glue. Recall was initiated due to the rate of the failure.

Distribution pattern

US Distribution including the states of OH, NC, and TX

Code information

Model/Product Number: B30-102 Lot #416130 Exp. 12/2016

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0023-2015
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Biomedix, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0023-2015. Report a data issue: hello@regdatalab.com.