Z-0032-2026 Class II Ongoing

Recalled by Beckman Coulter Inc. — Brea, CA

Recall Details

Product Type
Devices
Report Date
October 8, 2025
Initiation Date
August 4, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
116

Product Description

DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer

Reason for Recall

On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.

Distribution Pattern

Internationally distribution to the countries of Poland, India, Dominican Republic, Lebanon, Italy, Brazil, El Salvador, Saudi Arabia, Germany, France, Nepal, Canada, Chile, Spain, Malaysia, Switzerland, Morocco, Philippines, Curacao, Qatar, United Arab Emirates, Hungary, Czechia, Oman, New Zealand, Thailand, Indonesia, Algeria, United Kingdom of Great Britain and Northern Ireland, Namibia, Turkey, Republic of Korea

Code Information

REF/UDI-DI: C63522/14987666545089, C63521/14987666545072. Software Versions: SW 1.3.3 and 1.3.4