Z-0037-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 15, 2014
- Initiation Date
- July 9, 2014
- Termination Date
- April 24, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 73 cartons (1,460 clips)
Product Description
The Hulka Clip is a sterile packaged medical device that carries a two year expiration date. It is packaged two clips per procedural tray. Theses procedural trays are then sealed in a Tyvek/Mylar pouch and packaged in a cardboard shipping carton with ten pouches in each shipping carton.
Reason for Recall
Richard Wolf Medical Instrument Corporation is recalling Hulka Clip Tubal Occluding Device lot 51000232. The devices are being recalled as a part of an investigation into a biocompatibility test discrepancy where the recalled lot (lot 51000232) passed final release testing, but the subsequent lot failed biocompatibility testing on a component that was also used to manufacture lot 51000232.
Distribution Pattern
Worldwide distribution: US (Nationwide) including states of: (AK, AL, CO, CT, FL, GA, IA, IN, KY, LA, MI, MO, MS, NC, NY, OH, OR, PA, SC, TN, TX, VA, WA, WV; and country of: Germany.
Code Information
Product Code: 4986.09; Lot: 51000232