Z-0037-2026 Class II Ongoing

Recalled by Canon Medical System, USA, INC. — Tustin, CA

FDA device recall Z-0037-2026 was initiated by Canon Medical System, USA, INC. on Sep 2, 2025 and is designated Class II. Reason for recall: It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI… The recall status is ongoing. Affected quantity: 8.

Recall details

Product Type
Devices
Report Date
Oct 15, 2025
Initiation Date
Sep 2, 2025
Voluntary/Mandated
FDA Mandated
Product Quantity
8

Product description

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Reason for recall

It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.

Distribution pattern

U.S.

Code information

Serial Numbers: SZA2422001, SXA2432001, SUC23Z2014, BBF2392003, SUC2392013, SZA2422002, SXA2432002, SZA2442004.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0037-2026
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Canon Medical System, USA, INC.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0037-2026. Report a data issue: hello@regdatalab.com.