Z-0039-2014 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- November 6, 2013
- Initiation Date
- July 27, 2010
- Termination Date
- September 11, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Product Description
Xoran Minicat00158
Reason for Recall
The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha
Distribution Pattern
Worldwide Distribution - US nationwide
Code Information
minicat