Z-0059-2021 Class II Terminated

Recalled by AB ULAX — Motala, N/A

FDA device recall Z-0059-2021 was initiated by AB ULAX on Sep 4, 2020 and is designated Class II. Reason for recall: Firm has received 6 complaints that the Heat Moisture Exchanger (HME) disconnected from the closed respiratory circuit during use. If the HME separates, the ventilator circuit may not provide adequate… The recall status is terminated (terminated Feb 10, 2023). Affected quantity: 1,484,100 units.

Recall details

Product Type
Devices
Report Date
Oct 14, 2020
Initiation Date
Sep 4, 2020
Termination Date
Feb 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,484,100 units

Product description

1500 Flex Heat and Moisture Exchanger, Catalog Reference 150 - Product Usage: Heat Moisture Exchangers (HM Es) are designed to enhance humification of ventilator-dependent and non-ventilator dependent patients in critical care, home care and anesthesia.

Reason for recall

Firm has received 6 complaints that the Heat Moisture Exchanger (HME) disconnected from the closed respiratory circuit during use. If the HME separates, the ventilator circuit may not provide adequate ventilation to the patient, potentially resulting in lack of oxygen leading to hypoxia or death.

Distribution pattern

Worldwide distribution - US Nationwide distribution and the countries of Sweden, Taiwan, Hong Kong, and Spain.

Code information

SAP 109381602, Lots: 2018-09-25 2018-11-20 2018-12-10 2018-12-20 2019-01-03 2019-02-20 2019-03-20 2019-04-15 2019-05-30 2019-06-25 2019-08- 15 2019-09-05 2019- 11-05 2019-11-25 *** SAP 109381603, Lots 2019-12-20 2020-04-30 2020-05-15 2020-05-20 2020-06-15 2020-06-30

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0059-2021
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for AB ULAX? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0059-2021. Report a data issue: hello@regdatalab.com.