Z-0059-2021 Class II Terminated

Recalled by AB ULAX — Motala, N/A

Recall Details

Product Type
Devices
Report Date
October 14, 2020
Initiation Date
September 4, 2020
Termination Date
February 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,484,100 units

Product Description

1500 Flex Heat and Moisture Exchanger, Catalog Reference 150 - Product Usage: Heat Moisture Exchangers (HM Es) are designed to enhance humification of ventilator-dependent and non-ventilator dependent patients in critical care, home care and anesthesia.

Reason for Recall

Firm has received 6 complaints that the Heat Moisture Exchanger (HME) disconnected from the closed respiratory circuit during use. If the HME separates, the ventilator circuit may not provide adequate ventilation to the patient, potentially resulting in lack of oxygen leading to hypoxia or death.

Distribution Pattern

Worldwide distribution - US Nationwide distribution and the countries of Sweden, Taiwan, Hong Kong, and Spain.

Code Information

SAP 109381602, Lots: 2018-09-25 2018-11-20 2018-12-10 2018-12-20 2019-01-03 2019-02-20 2019-03-20 2019-04-15 2019-05-30 2019-06-25 2019-08- 15 2019-09-05 2019- 11-05 2019-11-25 *** SAP 109381603, Lots 2019-12-20 2020-04-30 2020-05-15 2020-05-20 2020-06-15 2020-06-30