Z-0069-2022 Class II Ongoing

Recalled by CardioQuip, LLC — College Station, TX

Recall Details

Product Type
Devices
Report Date
October 13, 2021
Initiation Date
June 24, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1380 units

Product Description

CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RMS, UDI: (01)00860000846127 used with heat exchange devices.

Reason for Recall

There is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result in contaminated water potentially aerosolized into the operating room during surgery which could result in patient infection.

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of TX, PA, FL, CA, HI, SC, OH, AK, NY, IL, AR, WI, MI, AL, KS, NJ, GA, AZ, KY, MS, TN, LA, DE, MA, VA, CO, NE, NC, NH, ME, WV, PA, MO, MN, IN, CT, NH, NV, UT and the country of Venezuela.

Code Information

All Serial Numbers