Z-0099-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 21, 2020
- Initiation Date
- September 2, 2020
- Termination Date
- September 28, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 107 calibrator kits
Product Description
GENTIAN Cystatin C Calibrator and Calibrator Kit, for the following models: 1. Gentian C E, Cystatin C Calibrator, 1 x 1 mL, REF 1012, UDI: (01)07090038460092 2. Gentian C E Calibrator Kit, 6 x 1 mL, REF 1051, UDI: (01)07090038460108 3. Gentian C E, BCI Ref: A52763, Cystatin C calibrator Kit, 6 x 1 mL, REF 1051, (01)17090038460037 4. Gentian C E, Cystatin C Calibrator, 1 x 1 mL, REF 8012 (Not distributed in US)
Reason for Recall
The firm has identified lots of calibrators that have shown to measure incorrect high values over time compared to the assigned values when released for distribution. This could result in higher Cystatin C values measured in patients.
Distribution Pattern
US: MN, VA, KS,
Code Information
Lot numbers #s: 1905415,1904404,1904403,1809403