Z-0106-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 24, 2018
- Initiation Date
- April 4, 2017
- Termination Date
- May 14, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3 kits
Product Description
"Free and Total Protein S ELISA Kit" Cat. No. 5292 For use in the quantitation of free and total Protein antigen in citrated human plasma Contains: 96 Protein S: Ag Microwells, 3 x 0.5mL ELISA Reference Plasma, 1x12 mL Protein S Conjugate Solution, 1 x 60mL Sample Diluent, 1 x 13mL Substrate, 1 x 15mL Stopping Solution, 1 x 30mL Phosphate Buffered Saline, 1 x 2mL Free Protein S Reagent FOR IN-VITRO DIAGNOSTIC USE LOT: 3-16- 5292 EXP: 2018-01-26 HELENA LABORATORIES BEAUMONT, TX Outer labeling of Substrate (brown bottle and cap): "Substrate 13 mL" FOR IN-VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES Beaumont, Tx "2161421 EXP: 2018-01-26" Outer Labeling of Protein S Conjugate bottle (clear bottle and cap, red liquid): "Protein S Conjugate 12 mL" FOR IN-VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES Beaumont, Tx "2161421 EXP: 2018-01-26"
Reason for Recall
Internal components of the Protein S ELISA Antigen Kit (Lot: 3-16-5292) were mislabeled. The Protein S Conjugate Solution is labeled as Substrate, and the Substrate is labeled as Protein S conjugate.
Distribution Pattern
International distribution to country of: Jordan
Code Information
3-16-5292; UDI: Primary DI - M52552920