Z-0107-2024 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Chaska, MN

Recall Details

Product Type
Devices
Report Date
October 18, 2023
Initiation Date
September 12, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
36 units

Product Description

DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic

Reason for Recall

The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the same rack and position. This will cause a delay of result to the sample currently presented to the system (Sample B) and may result in an erroneous test result for the prior sample (Sample A), which gets associated with the results of sample B.

Distribution Pattern

Worldwide - US Nationwide distribution in the state of Florida, and the countries of Croatia, Czechia, France, Germany, Ireland, Italy, Netherlands, New Zealand, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.

Code Information

UDI/DI 15099590732103, All DxI 9000 Access Immunoassay Analyzer running software version 1.15 or prior