Z-0113-2021 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 28, 2020
- Initiation Date
- August 31, 2020
- Termination Date
- June 18, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2990 units
Product Description
Langston Dual Lumen Catheter.
Reason for Recall
The inner lumen of the Langston catheter may separate from the hub during pressure injection.
Distribution Pattern
Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL ,GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ,NM ,NV ,NY ,OH ,OK,OR, PA ,PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. International Distribution: Austria, Australia, Belgium, Bulgaria, Canada, Cyprus, Czech Republic, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Mexico, Netherlands, Oman, Poland, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates and United Kingdom.
Code Information
Model Number: 5550, UDI 10841156100537 total 2990, Lots: 637973, 640204, 659555, 660597, 633668, 663767, 669593, 659856, 644096, 676528, 672481, 632704, 656792, 658273, 666755, 635885.