Z-0117-2016 Class II Terminated

Recalled by Natus Neurology Inc — Middleton, WI

Recall Details

Product Type
Devices
Report Date
October 21, 2015
Initiation Date
September 11, 2015
Termination Date
March 8, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
101 (88 US, 13 OUS)

Product Description

NicoletOne Software v5.94, Catalog/Part Number: 482-649600. The NicoletOne data acquisition and review system, either with or without synchronous digital video. The system is intended for medical purposes to record, measure, store and display full band (FbEEG) cerebral EEG and extracerebral activity for Clinical EEG, Electrocorticography (EcOG), Long Term Monitoring (LTM), Intensive Care Unit (ICU) monitoring and Polysomnography (PSG) Sleep studies. While the Nicolet Neurodiagnostic systems are capable of displaying signals, such as Sp02 and EKG, the system is NOT intended for monitoring such signals for the preservation of life, The Nicolet Neurodiagnostic systems are intended to acquire, analyze, and display data

Reason for Recall

When using the NicoletOne v5.94 software, after exiting the impedance check function and returning to the EEG screen, the impedance check signal remains active in waveform, obscuring the EEG signals.

Distribution Pattern

Worldwide distribution. US nationwide, INDIA, ROMANIA, DENMARK, JAPAN, AUSTRALIA, RUSSIA, BANGLADESH, SRI LANKA, BRUNEI, and UNITED ARAB EMIRATES.

Code Information

Serial numbers include TS150456M, TR150171M, TR150172M, 1TR150173M, TR150174M, TS150451M, TS150452M, TR150170M, UH150119M, TR150169M, UR150226M, UR150227M, UR150228M, UR150229M, TS150455M, TS150457M, TS150458M, TS150454M, TS150453M, UC150184M, PN151700M, SJ150272M, TT150134M, PN151699M, UR150230M, UR150231M, and SJ150271M.