Z-0117-2016 Class II Terminated
FDA device recall Z-0117-2016 was initiated by Natus Neurology Inc on Sep 11, 2015 and is designated Class II. Reason for recall: When using the NicoletOne v5.94 software, after exiting the impedance check function and returning to the EEG screen, the impedance check signal remains active in waveform, obscuring the EEG signals. The recall status is terminated (terminated Mar 8, 2016). Affected quantity: 101 (88 US, 13 OUS).
Recall details
- Product Type
- Devices
- Report Date
- Oct 21, 2015
- Initiation Date
- Sep 11, 2015
- Termination Date
- Mar 8, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 101 (88 US, 13 OUS)
Product description
Reason for recall
Distribution pattern
Worldwide distribution. US nationwide, INDIA, ROMANIA, DENMARK, JAPAN, AUSTRALIA, RUSSIA, BANGLADESH, SRI LANKA, BRUNEI, and UNITED ARAB EMIRATES.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0117-2016
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
Need the full FDA record for Natus Neurology Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude