Z-0119-2022 Class I Ongoing

Recalled by Windstone Medical Packaging, Inc. — Billings, MT

FDA device recall Z-0119-2022 was initiated by Windstone Medical Packaging, Inc. on Aug 19, 2021 and is designated Class I. Reason for recall: Monoject 0.9% Sodium Chloride Flush Syringe Manufactured by Cardinal Health and placed into kits. Potential for the plunger to draw back after air has been expelled and reintroduce air back into the… The recall status is ongoing. Affected quantity: 9,378.

Recall details

Product Type
Devices
Report Date
Nov 10, 2021
Initiation Date
Aug 19, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9,378

Product description

Monoject 0.9% Sodium Chloride, Flush Prefilled Syringe, 10 mL, REF: 8881570121 inside the following Aligned Medical Solutions Custom Convenience Kits: AMS-9041CP Leaderflex Insertion Kit with Ultrasound, AMS-9046CP-1 Insertion Tray-RX, AMS8939A Universal Procedure Pack w/Split Drape, AMS9957A Port Insertion Pack, and AMS12149 Procedure Pack

Reason for recall

Monoject 0.9% Sodium Chloride Flush Syringe Manufactured by Cardinal Health and placed into kits. Potential for the plunger to draw back after air has been expelled and reintroduce air back into the syringe.

Distribution pattern

FL and PA

Code information

Syringe Lots: All Lots manufactured and distributed between July 1, 2019 and July 2021. Kit Model/Lots: AMS-9041CP/ 136833; AMS-9046CP-1/ 137794; AMS8939A/ 140390, 141088, 141096, 141100, 141102, 146039, 146349, 147902, 147903, 148414, 148452, 148640, 149132, 149135, 149485, 141087, 141097, 145615, 146351, 146766, 146768, 146905, 146908, 148638, 149137, 149138, 149484, 149486, 141101, 148410, 148411, 148639, 149133, 141098, 146037, 148409, 148413, 148450, 149131, 149134, 149136, 148412, 148451, 148636, 141717 AMS9957A/ 147773; AMS12149/ 145718, 146897, 148612

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0119-2022
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0119-2022. Report a data issue: hello@regdatalab.com.