Z-0136-2017 Class II Terminated

Recalled by Hand Biomechanics Lab Inc — Sacramento, CA

Recall Details

Product Type
Devices
Report Date
October 26, 2016
Initiation Date
September 14, 2016
Termination Date
November 15, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
65 units

Product Description

Agee-WristJack¿ Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. This is a single use system used for a trauma related surgery The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. The non-perishable sterile system is packaged in an inner then outer tray and shipped in a cardboard shelf box. Orthopedic:The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures.

Reason for Recall

A number of Agee WristJack¿ Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the surgeon experiencing difficulty fine-tuning the length (red) and radial/ulnar alignment (blue) adjustments on the device.

Distribution Pattern

Nationwide Distribution

Code Information

Lot number: HBL-115-165, serial numbers 47460959 and 47460980; Lot number: HBL-116-014 - all serial numbers; Lot number : HBL-116-040 - 15 systems shipped prior to recall