Z-0136-2025 Class II Ongoing

Recalled by B-K Medical A/S — Herlev, N/A

Recall Details

Product Type
Devices
Report Date
October 30, 2024
Initiation Date
September 4, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1648 devices

Product Description

bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66, ultrasound system with battery option.

Reason for Recall

The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.

Distribution Pattern

Worldwide - US Nationwide - There was also government/military distribution and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, and United Kingdom.

Code Information

All serial numbers, GTIN 05704916000264.