Z-0178-2014 Class II Terminated

Recalled by SEPPIM SAS — SEES, N/A

FDA device recall Z-0178-2014 was initiated by SEPPIM SAS on Sep 4, 2013 and is designated Class II. Reason for recall: Calcium Arsenazo Reagent, reference CALA-0250, is recalled due to complaint of plasma specimens report high results than serum. The recall status is terminated (terminated Dec 12, 2013). Affected quantity: 47 kits in US and 48 kits in Sweden.

Recall details

Product Type
Devices
Report Date
Nov 13, 2013
Initiation Date
Sep 4, 2013
Termination Date
Dec 12, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
47 kits in US and 48 kits in Sweden

Product description

Calcium Arsenazo, CALA-0250, For the Quantitative Determination of Total Calcium in Human Serum. The kit label shows CALCIUM ARSENAZO, CALA-0250, R: 12x20 mL. The vial's label shows CALCIUM ARSENAZO, CALA-5220, 20 mL, . CALCIUM ARSENAZO is intended use for the quantitative in vitro diagnostic determination of total calcium in human serum and plasma on ELITech Clinical Systems Selectra Pro Series Analyzers. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). It is not intended for use in Point of Care setting.

Reason for recall

Calcium Arsenazo Reagent, reference CALA-0250, is recalled due to complaint of plasma specimens report high results than serum.

Distribution pattern

Worldwide Distribution - US including the states of KY and UT and the country of Sweden.

Code information

Lot numbers distributed in the US: Lot 11-2614, Expiration 07/2013. Lot 12-1002; Expiration 03/2014. Lot 12-2220, Expiration 08/2014. Lot numbers distributed in Sweden: 11-2129, 12-2616, 11-2769, 12-2975, 10-3036, 10-0454, 10-1874, 11-0176, 11-1248, 11-1690, 12-0140, 12-0938, 12-1383, 13-0004, and 13-0408

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0178-2014
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for SEPPIM SAS? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works