Z-0178-2014 Class II Terminated

Recalled by SEPPIM SAS — SEES, N/A

Recall Details

Product Type
Devices
Report Date
November 13, 2013
Initiation Date
September 4, 2013
Termination Date
December 12, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
47 kits in US and 48 kits in Sweden

Product Description

Calcium Arsenazo, CALA-0250, For the Quantitative Determination of Total Calcium in Human Serum. The kit label shows CALCIUM ARSENAZO, CALA-0250, R: 12x20 mL. The vial's label shows CALCIUM ARSENAZO, CALA-5220, 20 mL, . CALCIUM ARSENAZO is intended use for the quantitative in vitro diagnostic determination of total calcium in human serum and plasma on ELITech Clinical Systems Selectra Pro Series Analyzers. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). It is not intended for use in Point of Care setting.

Reason for Recall

Calcium Arsenazo Reagent, reference CALA-0250, is recalled due to complaint of plasma specimens report high results than serum.

Distribution Pattern

Worldwide Distribution - US including the states of KY and UT and the country of Sweden.

Code Information

Lot numbers distributed in the US: Lot 11-2614, Expiration 07/2013. Lot 12-1002; Expiration 03/2014. Lot 12-2220, Expiration 08/2014. Lot numbers distributed in Sweden: 11-2129, 12-2616, 11-2769, 12-2975, 10-3036, 10-0454, 10-1874, 11-0176, 11-1248, 11-1690, 12-0140, 12-0938, 12-1383, 13-0004, and 13-0408