Z-0184-2022 Class II Ongoing

Recalled by Karl Storz Endoscopy — El Segundo, CA

Recall Details

Product Type
Devices
Report Date
November 10, 2021
Initiation Date
September 10, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
532

Product Description

FIVE S 3.5x65, Flexible Intubation Videoscope, REF: 091361-01 (1 Unit), and 091361-06 (6 Units)

Reason for Recall

During a regular requalification review, affected flexible intubation video endoscopes did not achieve the required sterility assurance level. There is a risk that the patient may be exposed to a higher risk of infection.

Distribution Pattern

US: SC, NH, UT, LA, FL, TX, MA, MO, VA, IA, CA, IL, NC, VT, AL, GA, PA, TN, WI, NY, CO, AZ, NV, ME, MS, DC, IN, KY, MD, OH, MI, WA, KS, OR, AR, MN, NJ, NM, WV, CT. OUS: Canada

Code Information

All LOTs with remaining shelf life