Z-0203-2022 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- November 10, 2021
- Initiation Date
- October 11, 2021
- Termination Date
- January 16, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 567 (US)
Product Description
GM Helix Acqua Implant 3.5x10, Article No. 140.944
Reason for Recall
3.5x11.5mm implants were mislabeled as 3.5x10mm implants. In borderline cases where the bone quality is low and the bone ridge has a maximum indication for the implant height (10mm), if the incorrect implant is installed, it could result in damage to the nerve or adjacent bone structures/cavities.
Distribution Pattern
US Nationwide distribution.
Code Information
UDI 7899878024897 Lots GNW15 and GRE82