Z-0224-2015 Class II Terminated

Recalled by Nidek Inc — Fremont, CA

FDA device recall Z-0224-2015 was initiated by Nidek Inc on Oct 24, 2012 and is designated Class II. Reason for recall: When the MC-500 is turned on and coagulation is performed without changing the laser color, the delivery unit occasionally fails to emit the laser, or the power output is lower than the set program. The recall status is terminated (terminated Nov 17, 2014). Affected quantity: 21affected devices.

Recall details

Product Type
Devices
Report Date
Nov 26, 2014
Initiation Date
Oct 24, 2012
Termination Date
Nov 17, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21affected devices

Product description

Nidek MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic Device. The Nidek Multicolor Laser photocoagulator Model MC-500 is indicated for use in retinal photocoagulation for treatment of ocular fundus diseases.

Reason for recall

When the MC-500 is turned on and coagulation is performed without changing the laser color, the delivery unit occasionally fails to emit the laser, or the power output is lower than the set program.

Distribution pattern

US Distribution in the states of: NJ, SC, IL, MI, NY, CA, TX, OH, IN, UT, and MN.

Code information

Model MC-500 Serial numbers: 50277 50310 50330 50331 50357 50358 50383 50408 50417 50418 50425 50426 50439 50440 50445 50446 50471 50514 50311 50359 50382,

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0224-2015
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Nidek Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0224-2015. Report a data issue: hello@regdatalab.com.