Z-0229-2023 Class II Ongoing

Recalled by DeVilbiss Healthcare LLC — Somerset, PA

Recall Details

Product Type
Devices
Report Date
November 23, 2022
Initiation Date
September 19, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
48 units

Product Description

125 series iGO2 Portable Oxygen Concentrator-indicated for the administration of supplemental oxygen. Models: 125D, 125D-XB

Reason for Recall

An unapproved adhesive used to manufacture the accumulator tank sub assembly (a part of the patient gas pathway)

Distribution Pattern

US Distribution in states of: CA, FL, OH, MI, MN, NY, PA, and TX; and (Foreign) country of: Canada.

Code Information

UDI: 00885304022237 S/N: New Sales D22829087DS D22829097DS D22829092DS D22829086DS D22829156DS D22901205DS D22901207DS D22829099DS D22829161DS D22901203DS D22829157DS D22829159DS D22901204DS D22901201DS D22829096DS D22826095DS D22826079DS D22826091DS D22826086DS S/N: Repair D22829087DS D22829097DS D22829092DS D22829086DS D22829156DS D22901205DS D22901207DS D22829099DS D22829161DS D22901203DS D22829157DS D22829159DS D22901204DS D22901201DS D22829096DS D22826095DS D22826079DS D22826091DS D22826086DS