Z-0234-2024 Class I Ongoing

Recalled by Unomedical A/S — LEJRE, N/A

Recall Details

Product Type
Devices
Report Date
November 29, 2023
Initiation Date
October 11, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3362 units

Product Description

VariSoft infusion set, Model Number 1002830

Reason for Recall

The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.

Distribution Pattern

US Nationwide distribution in the state of CA.

Code Information

GTIN 05705244018792, Serial Numbers: 5388368; GTIN 05705244018785, Serial Numbers: 5388368