Z-0234-2024 Class I Ongoing

Recalled by Unomedical A/S — LEJRE, N/A

FDA device recall Z-0234-2024 was initiated by Unomedical A/S on Oct 11, 2023 and is designated Class I. Reason for recall: The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming… The recall status is ongoing. Affected quantity: 3362 units.

Recall details

Product Type
Devices
Report Date
Nov 29, 2023
Initiation Date
Oct 11, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3362 units

Product description

VariSoft infusion set, Model Number 1002830

Reason for recall

The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.

Distribution pattern

US Nationwide distribution in the state of CA.

Code information

GTIN 05705244018792, Serial Numbers: 5388368; GTIN 05705244018785, Serial Numbers: 5388368

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0234-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0234-2024. Report a data issue: hello@regdatalab.com.