Z-0238-2022 Class II Ongoing

Recalled by Philips Healthcare — Andover, MA

Recall Details

Product Type
Devices
Report Date
December 1, 2021
Initiation Date
September 24, 2021
Termination Date
N/A
Voluntary/Mandated
FDA Mandated
Product Quantity
226 units

Product Description

Philips Azurion systems with software releases 2.1(L1) and 2.1(L2)

Reason for Recall

3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging. Before starting a 3D-RA scan, the detector of the Philips Azurion system must be positioned in portrait or landscape orientation with an angle of 0 or +/-90 degrees relative to the 3D scan direction. Philips has discovered that Azurion releases 2.1(L1) and 2.1(L2) allow creation of a 3D-RA scan with a deviating detector orientation of up to 3 degrees from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees), relative to the 3D scan direction. 3D-RA scans with a deviating detector orientation of ore than 1.0 degree from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees) cannot be reconstructed with the 3D-RA reconstruction software.

Distribution Pattern

US Nationwide distribution

Code Information

Azurion 2.1(L1) and 2.1(L2)