Z-0249-2023 Class II Ongoing

Recalled by 3M Healthcare Business — Saint Paul, MN

Recall Details

Product Type
Devices
Report Date
November 30, 2022
Initiation Date
October 6, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
248,750 units

Product Description

3M Steri-Drape, Small Towel Drape, REF 1000NSD, general surgery drape

Reason for Recall

During a recent investigation, 3M confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Additionally, 3M has observed an increase in reported adhesive related skin injuries for these affected lots.

Distribution Pattern

Worldwide distribution.

Code Information

UDI/DI (01)50707387445701, Lot Numbers: 33HCR7, 33HD4C, 33J4J6, 33J4NF, 33JRKR, 33JRYN, 33JT7E, 33K6D6, 33KAH8, 33KAMP, 33KAY7, 33KC8K