Z-0252-2016 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- November 18, 2015
- Initiation Date
- September 8, 2015
- Termination Date
- January 11, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 151 units
Product Description
Smooth Muscle Action (SMA) Immunohistochemistry reagent. Catalog numbers: CME305A-0.1ml; CME305B - 0.5 ml, PME305AA - 6.0 ml prediluted and ready to use. They are the same product with the same intended use. Chemistry - IVD Smooth Muscle Actin (SMA) antibody recognizes the alpha-smooth muscle isoform of actin. This antibody recognizes the alpha-smooth muscle isoform of actin. This MAb is reportedly useful for identifying tumors arising from smooth muscle and myoepithelial cells.
Reason for Recall
Certain catalog numbers and lots of Smooth Muscle Actin may stain some cell types that are typically expected to be negative for SMA.
Distribution Pattern
Worldwide Distribution - US including CO, FL, OH, SC, AZ, AL, FL, CT, NJ, CA, DC, KY, NC, TX, ID, MA, MD, OR, LA, PA, KS, NY, UT, GA and Internationally to Spain, Sweden, Switzerland, Hungary, Italy, Germany, Turkey, Canada, India, Bolivia, Mexico, Indonesia and Australia.
Code Information
Catalog numbers: CME305A & CME305B: lots 110112, Expiry 11/2015;0 42213, expiry 4/2016; 101013, expiry 10/2016; 121813, expiry 12/2016; 070214, expiry 7/2017; 020215, expiry 2/2018. Catalog number: PME305AA, lots 070513, expiry 07/2016; 032114, expiry 3/2017, 030315, expiry 3/2018.