Z-0276-2022 Class II Ongoing

Recalled by Instrumentation Laboratory — Bedford, MA

Recall Details

Product Type
Devices
Report Date
December 1, 2021
Initiation Date
October 13, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5828 kits (310 US)

Product Description

HemosIL ReadiPlasTin (20 mL), Part number 0020301400. An in vitro diagnostic thromboplastin reagent.

Reason for Recall

The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential risk of reporting an erroneous result.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, IA, IL, KS, ME, MI, NC, NY, OK, SC, and VA. The countries of Canada, Chile, China, Costa Rica, Italy, Japan, and South Korea. ***Updated 3/3/22*** Expansion includes states of IN, MN, and NC.

Code Information

Lot N0705526 UDI: 08426950632887 ***Updated 3/3/22*** Firm has expanded recall to include the following lot: N0806224