Z-0288-2013 Class II Terminated

Recalled by Trivascular, Inc — Santa Rosa, CA

FDA device recall Z-0288-2013 was initiated by Trivascular, Inc on Oct 19, 2012 and is designated Class II. Reason for recall: Selected lots of the Fill Polymer Kit (European Model Number: TVFP14- B) in European distributor inventory only that potentially contain leaking stopcocks due to possible cracks in the stopcock… The recall status is terminated (terminated Dec 17, 2012). Affected quantity: 41 units.

Recall details

Product Type
Devices
Report Date
Nov 21, 2012
Initiation Date
Oct 19, 2012
Termination Date
Dec 17, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
41 units

Product description

TriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The TriVascular Ovation Abdominal Stent Graft System is indicated in subjects diagnosed with an aneurysm in the abdominal aorta having vascular morphology suitable for endovascular repair, including: (1) Adequate iliac/femoral access compatible with vascular access techniques, devices, and/or accessories, (2) Non-aneurysmal proximal aortic neck: - with a length of at least 7 mm proximal to the aneurysm, - with an inner wall diameter of no less than 16 mm and no greater than 30 mm, and - with an aortic angle of d 60 degrees if proximal neck is e 10 mm and d 45 degrees if proximal neck is < 10 mm, (3) Non-aneurysmal distal iliac landing zone: - with a length of at least 10 mm, - with an inner wall diameter of no less than 8 mm and no greater than 20 mm

Reason for recall

Selected lots of the Fill Polymer Kit (European Model Number: TVFP14- B) in European distributor inventory only that potentially contain leaking stopcocks due to possible cracks in the stopcock component. TriVascular indicated that these Fill Polymer Kits are for use with the Ovation or Ovation Prime Abdominal Stent Graft Systems that may display leaks upon initiation of mix but prior to use in

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of Belgium, Cyprus, Italy, Spain, Switzerland and Turkey. TriVascular stated that no purchased devices are in hospital inventory in Europe or the U.S.

Code information

The following lot numbers are in distributor inventory in Europe: FF121211-04, FF010912-01, FF013112-01, FF020912-01, FF022212-03, FF022912-01, FF040412-01, FF051012-01

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0288-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Trivascular, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Trivascular, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Trivascular, Inc →

Source: openFDA record for Z-0288-2013. Report a data issue: hello@regdatalab.com.