Z-0290-2021 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- November 4, 2020
- Initiation Date
- September 9, 2020
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13860 units
Product Description
GORE TAG CONFORMABLE Thoracic Stent Graft with ACTIVE CONTROL System
Reason for Recall
Due to secondary deployment fiber being attached to secondary deployment handle, completion of deployment via the deployment line access hatch was not possible
Distribution Pattern
U.S. - Nationwide O.U.S. - Australia, Austria, Belgium, Croatia, Cypress, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Lebanon, Luxembourg, Monaco, Netherlands, Norway, New Zealand, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom
Code Information
CATALOG NUMBERS TGM212110X* TGM212115X* TGM212120X* TGM262610X* TGM262615X* TGM262620X* TGM282810X* TGM282815X* TGM282820X* TGM313110X* TGM313115X* TGM313120X* TGM343410X* TGM343415X* TGM343420X* TGM373710X* TGM373715X* TGM373720X* TGM404010X* TGM404015X* TGM404020X* TGM454510X* TGM454515X* TGM454520X* TGM262110X* TGM312610X* TGMR313110X* TGMR313115X* TGMR313120X* TGMR312610X* TGMR373710X* TGMR373715X* TGMR373720X* TGMR404010X* TGMR404015X* TGMR404020X* *X is a placeholder for enumeration code: E=EMEA, Australia, New Zealand; J=Japan All lot numbers