Z-0309-2021 Class II Terminated

Recalled by Battelle Memorial Institute — Columbus, OH

Recall Details

Product Type
Devices
Report Date
November 4, 2020
Initiation Date
August 13, 2020
Termination Date
February 3, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
898

Product Description

Critical Care Decontamination System (CCDS) Compatible N95 respirators - Product Usage: Used in decontaminating compatible N95 respirators2 for multiple-user3 reuse by healthcare personnel (HCP)4 to prevent exposure to pathogenic biological airborne particulates when there are insufficient supplies of Filtering Facepiece Respirators (FFRs) resulting from the Coronavirus Disease 2019 (COVID-19) pandemic.

Reason for Recall

Masks processed at one site were not maintained at levels of condensation during a portion of the decontamination cycle.

Distribution Pattern

US Nationwide distribution including in the states of CT, MA, RI.

Code Information

All decontaminated compatible N95 respirators distributed from Providence, RI between 07/01/2020 and 08/12/2020.