Z-0314-2017 Class II Terminated

Recalled by BBI SOLUTIONS OEM LTD — Dundee

Recall Details

Product Type
Devices
Report Date
November 2, 2016
Initiation Date
June 1, 2016
Termination Date
June 14, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
420 units (183 units quarantined)

Product Description

SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin degradation products containing the cross-linked D-diner site in human whole blood

Reason for Recall

A component of the kit is showing a sensitivity performance detected in stability testing at 15 months of an 18-month shelf life. Qualitative results may be affected. False negative results are possible which may incorrectly influence the clinical diagnostic decision.

Distribution Pattern

US distribution to CT and IN

Code Information

Lot Numbers containing test reagent lot # 1012-34437): BU183A, BU183B, BU184A, BU185A, BU185B Expiration Date: 31 JUL 2016