FDA device recall Z-0329-2025 was initiated by WISSNER-BOSSERHOFF GMBH on July 11, 2022 and is designated Class II. Reason for recall: When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can… The recall status is ongoing. Affected quantity: 77 units.
Recall Details
- Product Type
- Devices
- Report Date
- November 20, 2024
- Initiation Date
- July 11, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 77 units
Product Description
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198
Reason for Recall
When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.
Distribution Pattern
US Nationwide distribution in the states of NC.
Code Information
UDI/DI 4050621100401, 19-0990 UDI/DI 4050621100012, 19-1062 UDI/DI 4050621100432, 19-1152 UDI/DI 4050621100449, 19-1119 UDI/DI 4050621100470, 19-1155 UDI/DI 4050621100487, 19-1156