Z-0332-2023 Class I Terminated

Recalled by Remel Inc — Lenexa, KS

Recall Details

Product Type
Devices
Report Date
December 14, 2022
Initiation Date
October 20, 2022
Termination Date
September 30, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1647 units

Product Description

Thermo SCIENTIFIC, Sensititre CMC7AFLF, Gram Negative IVD AST

Reason for Recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Distribution Pattern

US Nationwide Distribution

Code Information

Lot Numbers: B2281, B2151A, B2034, B1453