Z-0347-2023 Class II Terminated

Recalled by Greer Laboratories, Inc. — Lenoir, NC

FDA device recall Z-0347-2023 was initiated by Greer Laboratories, Inc. on Oct 25, 2022 and is designated Class II. Reason for recall: The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by… The recall status is terminated (terminated Feb 11, 2026). Affected quantity: 218 units.

Recall details

Product Type
Devices
Report Date
Dec 7, 2022
Initiation Date
Oct 25, 2022
Termination Date
Feb 11, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
218 units

Product description

GREER M. FAENI, 0.25 mL 20 MG/ML, Item: K6, Analyte Specific Reagent, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.

Reason for recall

The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency. The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the lot was misinterpreted.

Distribution pattern

US Distribution to states of: CA, CO, IA, KY, MA, MD, MO, NC, PA, UT, VA, and International to countries of: Canada, Denmark

Code information

Lot Number 389591, exp 6/12/2023

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0347-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Greer Laboratories, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Greer Laboratories, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Greer Laboratories, Inc. →

Source: openFDA record for Z-0347-2023. Report a data issue: hello@regdatalab.com.