Z-0353-2018 Class II Terminated

Recalled by Zoll Manufacturing Corp. — Pittsburgh, PA

Recall Details

Product Type
Devices
Report Date
January 24, 2018
Initiation Date
September 12, 2017
Termination Date
August 10, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
33,670 units

Product Description

LifeVest Wearable Defibrillator Model 4000, Product Number 10A0988-A01. The LifeVest system is indicated for patients 18 years of age and older who are at risk for sudden cardiac arrest and are not candidates for or refuse an implantable defibrillator.

Reason for Recall

Incorrect service code for properly catching critical defects during self-check. Potential for defibrillation shock failure

Distribution Pattern

US Nationwide

Code Information

All Serial Numbers