Z-0366-2026 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics, Inc. — East Walpole, MA

Recall Details

Product Type
Devices
Report Date
November 5, 2025
Initiation Date
September 22, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3318 units (169 US, 3149 OUS)

Product Description

Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2) L2KPP6;

Reason for Recall

The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AL, CA, DC, FL, ID, IN, LA, MD, NC, NJ, NY, PA, SC and the countries of Argentina, Australia, Bolivia, Bosnia Herzeg, Brazil, Bulgaria, Canada, Chile, Colombia, Cyprus, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, Hungary, India, Iran, Israel, Italy, Kosovo, Kuwait, Latvia, Lebanon, Macedonia, Mexico, Nepal, Netherlands, P.R. China, Pakistan, Palestinian Ter, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Serbia, Slovakia, Slovenia, South Africa, Spain, Taiwan, Tunisia, Turkey, Turkmenistan, U.A.E., Ukraine, Uruguay, Uzbekistan.

Code Information

1. Siemens Material Number (SMN): 10381441; Catalog Number: L2KPP2; UDI-DI: 00630414961910; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389; 2. Siemens Material Number (SMN): 10381442; Catalog Number: L2KPP6; UDI-DI: 00630414961927; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389;