Z-0388-2019 Class II Terminated

Recalled by GE Healthcare Biosciences — Westborough, MA

FDA device recall Z-0388-2019 was initiated by GE Healthcare Biosciences on Sep 10, 2018 and is designated Class II. Reason for recall: Non-conformance with the DeltaVision OMX SR, the interlocks are not wired in a redundant fashion to make the interlock system single-fault tolerant. The recall status is terminated (terminated May 12, 2020). Affected quantity: 39.

Recall details

Product Type
Devices
Report Date
Nov 28, 2018
Initiation Date
Sep 10, 2018
Termination Date
May 12, 2020
Voluntary/Mandated
FDA Mandated
Product Quantity
39

Product description

DeltaVision OMX SR Imaging System Model Number: 29115476 The DeltaVision OMX SR Imaging System is a piece of laboratory equipment for super resolution imaging offluorescently-la beling biological samples. Techniques such as 3D Structured Illumination, Total Internal Reflection Fluorescence and localization microscopy may be used to examine a variety of samples over long periods of time.

Reason for recall

Non-conformance with the DeltaVision OMX SR, the interlocks are not wired in a redundant fashion to make the interlock system single-fault tolerant.

Distribution pattern

Worldwide - US Nationwide Distribution

Code information

System Serial number OM20072 - OM40122

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0388-2019
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for GE Healthcare Biosciences? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for GE Healthcare Biosciences: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about GE Healthcare Biosciences →

Source: openFDA record for Z-0388-2019. Report a data issue: hello@regdatalab.com.