Z-0398-2021 Class II Terminated

Recalled by Gordian Surgical — Misgav, N/A

FDA device recall Z-0398-2021 was initiated by Gordian Surgical on Sep 22, 2020 and is designated Class II. Reason for recall: Endotoxin test results found above the acceptable levels in LAL test (above 20EU/device). The recall status is terminated (terminated Feb 11, 2021). Affected quantity: 300 total devices.

Recall details

Product Type
Devices
Report Date
Nov 11, 2020
Initiation Date
Sep 22, 2020
Termination Date
Feb 11, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
300 total devices

Product description

TroClose 1200 - Product Usage: intended for use in a variety of gynecologic, general and urologic endoscopic procedures, to create and maintain a port of entry and to facilitate the delivery of absorbable sutures and anchors through soft tissues of the body during endoscopic / laparoscopic surgery. The trocar may be used with or without visualization for primary and secondary insertions. It is comprised of a single patient use, bladeless Obturator designed to allow penetration and positioning at the required site within the abdomen, and several single use Cannulas, which serve as working channels for the procedure. The Cannulas are pre-loaded with two synthetic absorbable sutures, each attached to a synthetic absorbable anchor. The deployment of the anchors and sutures is achieved by utilizing a mechanism within the Obturator. The Cannula is 115 mm long and is compatible for use with laparoscopic devices of 5 - 12 mm outer diameter.

Reason for recall

Endotoxin test results found above the acceptable levels in LAL test (above 20EU/device).

Distribution pattern

US Nationwide distribution including in the states of Texas and Michigan.

Code information

Catalog Number: GSTC1200 Lot Numbers: AS200009/3, Exp 14 FEB 2022 AS200012/4, Exp 23 MAR 2022

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0398-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Gordian Surgical? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Gordian Surgical: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Gordian Surgical →

Source: openFDA record for Z-0398-2021. Report a data issue: hello@regdatalab.com.