Z-0416-2026 Class II Ongoing

Recalled by Spacelabs Healthcare, Ltd. — Hertford, N/A

FDA device recall Z-0416-2026 was initiated by Spacelabs Healthcare, Ltd. on Sep 16, 2025 and is designated Class II. Reason for recall: Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality… The recall status is ongoing. Affected quantity: 142.

Recall details

Product Type
Devices
Report Date
Nov 12, 2025
Initiation Date
Sep 16, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
142

Product description

Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient and facility information.

Reason for recall

Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record

Distribution pattern

Worldwide - US Nationwide distribution in the state of MN and the countries of CA, DE, FR, GB, IT, NC.

Code information

Catalog Number: 92810 UDI-DI code; (01)10841522125751 Software version: 11.6.0

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0416-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.