Z-0419-2013 Class II Terminated

Recalled by King Systems Corp. — Noblesville, IN

FDA device recall Z-0419-2013 was initiated by King Systems Corp. on Nov 2, 2012 and is designated Class II. Reason for recall: The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change… The recall status is terminated (terminated May 28, 2013). Affected quantity: 263 cases of product. (2,630 devices).

Recall details

Product Type
Devices
Report Date
Nov 28, 2012
Initiation Date
Nov 2, 2012
Termination Date
May 28, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
263 cases of product. (2,630 devices)

Product description

KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 3 / Child Silicone King LAD¿ Patient Size: 30-50kg Product Usage: Laryngeal airway

Reason for recall

The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO

Distribution pattern

US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC

Code information

REF: LAD-303, Lot #K120618

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0419-2013
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for King Systems Corp.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0419-2013. Report a data issue: hello@regdatalab.com.