Z-0425-2023 Class II Ongoing

Recalled by Qvella Corporation — Richmond Hill, N/A

FDA device recall Z-0425-2023 was initiated by Qvella Corporation on Nov 2, 2022 and is designated Class II. Reason for recall: Device exceeds the limitations of a Class I 510(k)-exempt device so the user manual, instructions for use, and promotional materials are being corrected to remove all references to downstream… The recall status is ongoing. Affected quantity: 1868.

Recall details

Product Type
Devices
Report Date
Dec 14, 2022
Initiation Date
Nov 2, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1868

Product description

The FAST System and the FAST PBC Prep Cartridge together are a closed, rapid and automated sample preparation system for isolating bacterial cells directly from a positive blood culture (PBC). The cartridge and instrument isolate and concentrate bacterial cells directly from PBC by removing blood, cellular and culture media matrix components from the sample, thereby reducing the sample volume considerably while maintaining bacterial cell viability. The resulting material is a purified microbial suspension or Liquid Colony for use in downstream in vitro diagnostics.

Reason for recall

Device exceeds the limitations of a Class I 510(k)-exempt device so the user manual, instructions for use, and promotional materials are being corrected to remove all references to downstream applications.

Distribution pattern

US: CA, NY, AL, UT

Code information

TSP-UM-001, Revisions: 2 and 3; MKT-0025, Revisions: 2 and 3; TSP-PI-001, Revisions: 2 and 3

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0425-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Qvella Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Qvella Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Qvella Corporation →

Source: openFDA record for Z-0425-2023. Report a data issue: hello@regdatalab.com.