Z-0454-2016 Class II Terminated

Recalled by DePuy Spine, Inc. — Raynham, MA

Recall Details

Product Type
Devices
Report Date
December 30, 2015
Initiation Date
November 20, 2015
Termination Date
July 3, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 devices

Product Description

Aegis Torque Handle CATALOG NO. : 2871-10-800. A spinal surgical instrument used on the AEGIS Anterior Lumbar Plate System.

Reason for Recall

Manufacturing error within the Torque Handle may prohibit some handles from being able to fully assemble with the tightener shaft. The design of the inspection gage does not account for the additional shift length

Distribution Pattern

US Distribution to the states of : NC, MI, CO, MD, FL, ID, TX, CA and NY.

Code Information

LOT NUMBER: GB0906 GB0707 GB0907 GB58815 GB59244 GB61134