Z-0455-2022 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- January 12, 2022
- Initiation Date
- November 8, 2021
- Termination Date
- March 29, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 376 kits
Product Description
QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223
Reason for Recall
The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur.
Distribution Pattern
Domestic distribution to AL, DC, KY, MD, ME, NJ, OK< PA, and TX. No foreign distribution.
Code Information
LOT 210209 UDI (01)14053228038846(17)211127(10)210209(24)691223 Serial Numbers: 412092551 412092559 412092552 412092560 412092553 412092561 412092554 412092563 412092555 412092564 412092556 412092565 412092557 412092566 412092558 412092567